G005723
LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
column: Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)
mobile phase: [A] 10 mM ammonium acetate in water, pH 4.2; [B] 10 mM ammonium acetate in 90:10 (v/v) acetonitrile:water, pH 4.7
gradient: held at 20% B for 0.1 min; 20 to 80% B in 2.9 min; 80 to 20% B in 0.01 min; held at 20% B for 0.99 min
flow rate: 0.3 mL/min
column temp.: 35 °C
detector: TOF, ESI(+), m/z 100-2500
injection: 3 μL
sample: 1 μL/mL in 10mM ammonium formate, 80:20, acetonitrile:water
suitability
application for HPLC
application(s)
life science and biopharma
Analysis Note
Legal Information
Analyte
analytical column
mobile phase component
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service