LC/MS/MS Analysis of Carbendazim in Orange Juice (Spiked) on Ascentis® Express C18 after Supel QuE PSA/C18 QuEChERS Cleanup

LC/MS/MS Analysis of Carbendazim in Orange Juice (Spiked) on Ascentis® Express C18 after Supel™ QuE PSA/C18 QuEChERS Cleanup application for HPLC, application for SPE

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express C18, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

SPE tube or plate

Product No.
Description
Pricing

Supel QuE

Citrate Extraction Tube, pk of 50, suitable for EN 15662

related product

Product No.
Description
Pricing

Supel QuE

PSA Tube, pack of 100, AOAC® 2007.01

Supel QuE

PSA/C18 Tube, pack of 100, AOAC® 2007.01

Supel QuE

Non-Buffered Tube 1, pk of 50

CONDITIONS

sample preparation

Dispersive (QuEChERS)

sample/matrix

orange juice spiked with carbendazim at 1 ppb

extraction tube

Supel QuE Citrate (EN) Tube (55227-U)

extraction process

add 10 mL of orange juice with pulp to a 50 mL empty extraction tube (55248-U); add carbendazim at 1 ppb; add 10 mL acetonitrile; (shake for 1 minute; add contents of citrate extraction tube (55227-U); shake for 1 min; centrifuge at 3200 rpm for 5 minutes)

clean-up tube

Supel-Q PSA/C18 (55288-U)

clean-up process

transfer 0.7 mL of the acetonitrile layer to the PSA/C18 cleanup tube (55288-U); shake 1 minute; (centrifuge for 3 minutes at 3400 rpm; remove 0.25 mL of the supernatant; mix with 0.25 mL water for LC-MS analysis)

column

Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)

mobile phase

[A] 10 mM ammonium acetate in water; [B] 10 mM ammonium acetate in methanol

gradient

0-1 min: 30% B; 1.5-3.5 min: 100% B; 3.5-7 min: 30% B

flow rate

0-1 min: 0.3 mL/min; 1.5-7 min: 0.5 mL/min

column temp.

30 °C

detector

MS, ESI(+), MRM, m/z 192/160, 192/132

injection

5 μL

Description

Analysis Note

Simple and fast extraction and analytical method were developed for analysis of fungicide carbendazim in orange juice samples. QuEChERS extraction method was used with PSA/C18 cleanup. Ascentis Express C18 column chromatography was coupled to MS/MS detection with limits of detection at 0,1 ppb.